STANDARD OPERATING PROCEDURE Operating and Calibration Procedure for TOC Analyzer
|
STANDARD OPERATING PROCEDURE |
|
Department: Quality Control
|
SOP No.: QC/
|
|
Operating and Calibration Procedure for
TOC Analyzer
|
Effective
Date:
|
|
Review
Date:
|
||
Page
No.:
|
1.0
OBJECTIVE:
1.1
To
lay down a procedure for operation and calibration of TOC analyzer.
2.0
SCOPE:
2.1
This
SOP is applicable to the TOC Analyzer located in Quality Control department at Pharma Private Limited.
ID.No:
3.0
RESPONSIBILITY:
3.1
QC Chemist/Microbiologist – Responsible for operation
and calibration of the instrument.
3.2
Section In-charge - To
review the procedure and reports.
3.3
QC Head – Responsible for implementing the procedure.
3.4
QA Head – Overall compliance.
4.0
PROCEDURE:
4.1
Operating Procedure:
4.1.1
Switch
on the power and mains of the instrument and check the following before operating
the equipment.
4.1.2
The pressure of the zero air/nitrogen gas should be
within the range of 44 – 87 psi (If not turn the key cylinder and adjust the
pressure).
4.1.3
The level of water in the humidifier should be between
the lower level and the upper level marks. (if not, open the humidifier and
fill water up to the level).
4.1.4
The rinse water bottle is filled to above the 2L mark
on the bottle. The tip of the rinse tube reaches nearly to the bottom of the
rinse water bottle.
4.1.5
Check &fill
25% of phosphoric acid in the phosphoric acid bottle kept in left side of the TOC
analyzer (when ever required).
4.1.6
Check &fill 25% 2N hydrochloric acid in the
hydrochloric acid bottle and keep
it in the left side of the TOC analyzer (whenever required).
4.1.7
After completion
of above mentioned procedure press the power button on the front panel of
the instrument.
4.1.8
Then the initial display shows initializing on the
display.
4.1.9
Setting of carrier gas flow rate:
4.1.9.1
Open the front door of the instrument
4.1.9.2
Turn the carrier gas flow adjustment knob, adjust the
flow rate to 150mL/min.
4.1.9.3
After all adjustments check background monitor by
pressing F3 key, and check the following parameters to be used for start
the analysis.
4.1.9.4
Check the air pressure about 200 kilopascal and
flow rate is about 150mL/min. A small window comprising of the Background
Monitor will appear on the screen
4.1.9.5
The Background Monitor consist of the following
parameters
4.1.9.5.1
Furnace: 680ºc
4.1.9.5.2
Dehumidifier: 1.4
4.1.9.5.3
Baseline Position: OK
4.1.9.5.4
Baseline
Fluctuation: OK
4.1.9.5.5
Baseline
Noise: OK
4.1.9.6
After
attaining the all parameters, green light will flash to indicate the system
ready for analysis.
4.1.9.7
Then
press return key (F1) to go back to main screen.
4.1.10
Measuring of sample:
4.1.10.1
Press ASI key followed by Measure Sample key.
4.1.10.2
The measurement setting screen is displayed. Then
enter required values as given below.
4.1.10.3
Sample
name: Enter the sample name (max 16 characters) by using numeric
keys (use shift button for alphabets), then press enter key.
4.1.10.4
Vial
number: Move the cursor from the sample name to vial number and then
enter the vial no 1 to 8 and press enter key.
4.1.10.5
Press
F5 key the display shows the following parameters and use ↓ key to move the cursor to
each parameter.
4.1.10.5.1 Calibration curve: Sets the calibration curve to be used (from two point calibration) and
press enter key.
4.1.10.5.2 Minimum Number of injections: Enter the Minimum number of sample
analysis injections (3) to be performed and press enter key.
4.1.10.5.3 Maximum number of injections: Enter the maximum number of sample injections (5) to be
performed.
4.1.10.5.4 Standard deviation: Enter the standard deviation value
(0.5%) and press enter key.
4.1.10.5.5 Coefficient variation:
enter the coefficient variation value (3%) and press enter key.
4.1.10.5.6 Dilution factor: Enter
the dilution factor (1) and press enter key.
4.1.10.5.7 Injection volume: Given
by system automatically (816 µL)
4.1.10.5.8 Acid addition ratio: Enter the acid addition ratio (0.5%)
and press enter key.
4.1.10.5.9 Sparge time: Enter the sparge time (0.5 min) and press
enter key.
4.1.10.6
After
setting the parameters press enter and then next key.
4.1.10.7
Keep
the sample in respective vial and press start.
4.1.10.8
After
analysis is completed, display shows Area and ppb.
4.1.10.9
After
completion of analysis process the press the stand by button then select power off
by using F5 button .Then press Execute (F6 key). Now the system asks for confirmation,
press Yes (F6 Key). The system automatically shut down. (App.20 min)
4.1.10.10
Then
put off main switch and main gas supply.
4.2
Calibration procedure:
4.2.1
Two point calibration: By Sucrose solution
4.2.1.1
Preparation of Sucrose Standard stock solution: Weigh accurately 11.9 mg of sucrose
which is previously dried at 105°C for 3 hours, and dissolve and diluted it to
100 mL with zero water (50ppm). Dilute 1 mL of this solution to 100 mL with
zero water. (0.50 mg of carbon per liter or 500 ppb).
4.2.1.2
Measure
standard response for blank and 500 ppb solution.
4.2.1.3
Blank:
Zero water (100ppb).
4.2.1.4
Calibration Procedure:
4.2.1.4.1
Press
data report (F6) from initial display.
4.2.1.4.2
Press
Cal. Curve list (F6).
4.2.1.4.3
Select
TC2 from Cal. Curve list by using UP and DOWN arrow
4.2.1.5
Press
edit key (F6)
4.2.1.6
Change
curve shift status from origin to none by pressing F4 Key.
4.2.1.7
Press
OK (F6)
4.2.1.8
Press
Return (F1)
4.2.1.9
Keep
the Blank (Zero Water) in Vial no.1
4.2.1.10
Keep the sucrose solution (500ppb) in vial no.2
4.2.1.11
Press
ASI button
4.2.1.12
Press
Cal Button
4.2.1.13
Press
2 followed by enter now display will show two calibration points.
4.2.1.14
Then
press Next button for two times.
4.2.1.15
Now
the display will show “Set Sample and Press START button”
4.2.1.16
Press
Start button to start the calibration.
4.2.1.17
After
completion of two point calibration change the shift option from none to origin by following the
point no. 4.2.1.4.1 to 4.2.1.4.7
4.2.2
Single point calibration (System
Suitability): By 1, 4- Benzoquinone solution
4.2.2.1
Preparation of 1, 4- Benzoquinone
solution: Dissolve
15 mg of 1, 4- Benzoquinone into 200 mL of zero water and diluted 1 mL of this
solution to 100 mL with zero water. (0.50 mg of carbon per liter or 500 ppb)
4.2.2.2
Measure standard response for 500 ppb solution.
4.2.2.3
Note: To
perform single point calibration, two point calibration shall be performed
first.
4.2.2.4
Calibration Procedure:
4.2.2.4.1
Follow
procedure as mentioned in two point calibration
4.2.1.1 to 4.2.1.17
4.2.2.4.2
Press ASI key followed by Measure Sample key.
4.2.2.4.3
Now the measurement setting screen will displayed.
Then enter required values as given below.
4.2.2.4.4
Sample
name: Enter the sample name (1, 4 Benzoquinone) by using numeric
keys (use shift button for alphabets), then press enter key.
4.2.2.4.5
Vial
number: Move the cursor from the sample name to vial number and then
enter the vial no 1 to 8 and press enter key.
4.2.2.4.6
Press
F5 key (NPOC) the display shows the following parameters and use ↓ key to move the cursor to each parameter.
4.2.2.4.7
Calibration curve: Sets the calibration curve to be used
(from two point calibration) and press enter key.
4.2.2.4.8
Minimum Number of injections: Enter the Minimum number of sample
analysis injections (3) to be performed and press enter key.
4.2.2.4.9
Maximum number of injections: Enter the maximum number of sample
injections (5) to be performed.
4.2.2.4.10 Standard deviation: Enter the standard deviation value
(0.5%) and press enter key.
4.2.2.4.11 Coefficient variation:
enter the coefficient variation value (3%) and press enter key.
4.2.2.4.12 Dilution factor: Enter
the dilution factor (1) and press enter key.
4.2.2.4.13 Injection volume: Given
by system automatically (816 µL).
4.2.2.4.14 Acid addition ratio: Enter the acid addition ratio (0.5%)
and press enter key.
4.2.2.4.15 Sparge time: Enter the sparge time (0.5 min) and
press enter key.
4.2.2.4.16 After setting the parameters press
enter and then Next key for two times.
4.2.2.4.17 Display will show “Set Sample and
Press START button”
4.2.2.4.18 Keep the sample in respective vial
and press start.
4.2.2.4.19 Press Start button to start the
calibration.
4.2.2.4.20 After analysis is completed, display
shows Area and ppb.
4.2.2.4.21 Formula for manual calculation of
recovery:
Rss - Rw
% Recovery = ---------------- X 100
Rs
– Rw
Where, Rss
= Mean area of 1, 4 Benzoquinone
Rw = Mean area of Zero Water
Rs = Mean area of Sucrose
4.2.3
Four point calibration: By Potassium hydrogen phthalate
solution
4.2.3.1
Preparation of PHP Standard stock solution: Weigh accurately 0.2125 g of
potassium hydrogen phthalate, dissolve and dilute with 100 mL zero water (1000
ppm carbon stock solution). Dilute 1 mL
of this solution to 100 mL with zero water. (10 ppm carbon stock solution)
4.2.3.2
Further dilutions like 250 ppb, 500 ppb, and 1000 ppb shall
be prepared though auto dilution by instrument. Here dilution factor is applied
automatically by the instrument that is 40, 20, & 10 respectively.
4.2.3.3
Measure
standard response for blank, 250 ppb, 500 ppb and 1000 ppb.
4.2.3.4
Blank:
Zero water (100ppb).
4.2.3.5
Calibration Procedure:
4.2.3.5.1
Press
data report (F6) from initial display.
4.2.3.5.2
Press
Cal. Curve list (F6).
4.2.3.5.3
Select
the 1000ppb calibration curve from Cal. Curve list by using UP and DOWN arrow
4.2.3.5.4
Press
edit key (F6)
4.2.3.5.5
Change
curve shift status from origin to none by pressing F4 Key.
4.2.3.5.6
Press
OK (F6)
4.2.3.5.7
Press
Return (F1)
4.2.3.5.8
Keep
Blank (zero water) in Vial no.1 and PHP solution in vial No. 2
4.2.3.5.9
Press
ASI button followed by Cal button
4.2.3.5.10 Select the calibration curve from
calibration curve list by pressing 1.
4.2.3.5.11 Then press “Next” button for two
times.
4.2.3.5.12 Display will show “Set Sample and
Press START button”
4.2.3.5.13 Press “Start” button to start the
calibration.
4.2.3.5.14 After analysis is completed, display
shows Area and ppb.
4.2.3.5.15 After completion of two point
calibration change the shift option from none to origin by following the point
no.4.2.3.5.1 to4.2.3.5.7.
4.2.3.6
Acceptance criteria:
4.2.3.6.1
%
Recovery of 1, 4 Benzoquinone is 85 % to 115 %.
4.2.3.6.2
Standard
deviation is NMT 0.5.
4.2.3.6.3
Coefficient
of variation is NMT 3.0 %.
4.2.3.6.4
Either
standard deviation or coefficient of variation shall meet.
4.3
After
usage of instrument, the usage details shall be entered in Annexure IV.
4.4
Frequency:
4.4.1
Two
point calibration: Weekly.
4.4.2
Single
point calibration (System Suitability): Monthly ± 3days.
4.4.3
Four
point calibration: Quarterly ± 7 days.
4.5
If
TOC analyzer is failed to meet an acceptance criteria of calibration, then out
of calibration shall be investigated as per current version of SOP QA/.
4.6
Precautions:
4.6.1
Do
not switch on the instrument without supply of zero air.
4.6.2
External
drain tubing connected to the waste port on the right side of the instrument
does not touch the surface of the liquid in the waste container.
4.6.3
Do
not switch off the instrument directly (except Emergency).
4.6.4
All
glassware used for TOC analysis shall be cleaned as per QC/
4.6.5
At
sample collection point wash bottles with purified water 2 to3 times by filling
top without leaving any air gap in sampling bottle.
4.6.6
Close
the cap and seal it with para film.
5.0
Abbreviations:
ppm: parts per million
|
NMT: Not More Than
|
ppb: parts per billion
|
QC: Quality Control
|
SOP: Standard Operating Procedure
|
IC: Inorganic Carbon
|
TOC: Total Organic Carbon
|
QA: Quality Assurance
|
NPOC: Non-Purgeable Organic Carbon
|
|
Comments
Post a Comment