STANDARD OPERATING PROCEDURE LAB EQUIPMENT DIGITAL AUTOMATIC TAP DENSITY APPARATUS
STANDARD
OPERATING PROCEDURE
SUBJECT :
LAB
EQUIPMENT DIGITAL AUTOMATIC TAP DENSITY APPARATUS
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
Version
No :
|
1.0 OBJECTIVE:
To
provide guidelines for Operation & Maintenance of Digital Automatic Tap
Density
Apparatus
2.0 SCOPE :
To
establish a procedure for an operation of
Digital Automatic Tap Density Apparatus
3.0 PROCEDURE:
3.1
Keep the instrument on a study
level and near to the electric connections
3.2 Fix
the connectors to the sockets as explained in the instruction Manual. Connect
the main
unit to 230m V 50 Hz Mains
supply.
3.3 Switch on the instrument, 4 line LCD
display becomes ON.
3.3
Press EDIT key and program the Test file for the sample explained in the
instruction
Manual.
3.4 Pass the sample through 1 mm screen.
3.5 The
cylinder must be absolutely dry from inside.
3.6 Weigh
the desire quantity of the sample and fill it into the cylinder
3.7 Carefully
level the sample without compacting and read the volume.
3.8 Place
the cylinder in its holder and mount it on the USP-I or USP –II port as per the
test
selected. Cover the unused port with the PVC Stopper supplied with the
instrument.
3.9
Ensure that the S.S. Rod of the holder is clean. Otherwise clean it
properly.
3.10
Apply a drop of oil if necessary.
3.11
Press START key and select the test file for the sample under test.
3.12
Program operator Name, Batch No. and sample ID
3.13 Enter the weight and Un-settled Volume of
the sample
3.14
Press start for initial tapping and observe that the cylinder is tapped
After initial tapping is over, read the volume of the sample and enter
it
3.15 Press START for initial tapping, after
final tapping is over read the volume of the sample and enter it.
3.16 The instrument calculates the percentage
Difference in volume between two successive tapping if the difference is less
than 2% then the instrument give the message. TEST OVER,
3.17 The instrument calculates Bulk Density,
Tapped Density, Compressibility Index and Hausner Ration of the sample.
3.18
If the difference is more than
2% then the instrument continues the test further tapping till the Percentage
difference in volume between two successive tapping is less than 2%
3.19
After completion of testing, clean the measuring cylinder &
instrument
3.20
Follow the above steps as and when instrument is used
4.0 FREQUENCY: As and when required.
5.0 RESPONSIBILITY: QA Executive
6.0 REVIEW GUIDE:.
7.0 ANNEXURE:
8.0 SOPs DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name
of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
X
|
x
|
Warehouse
|
X
|
x
|
General
|
X
|
x
|
Maintenance
|
X
|
x
|
Comments
Post a Comment