STANDARD OPERATING PROCEDURE LAB EQUIPMENT DIGITAL BOTTLE CAP TORQUE TESTER
STANDARD OPERATING PROCEDURE
|
SUBJECT :
LAB EQUIPMENT DIGITAL BOTTLE CAP TORQUE TESTER
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Operation & Maintenance
of Digital Bottle Cap Torque Tester.
2.0 POLICY:
To establish a procedure for an operation of Digital Bottle Cap Torque Tester.
3.0 PROCEDURE:
3.1. Ascertain
that the system is properly places on a hard & even surface. Plug the
adapter into a wall electric socket.
3.2. Place
the bottle between the grips and turn the screw till the bottle is rigidly held
in it.
3.3. Switch
on the system, wait for two minutes.
3.4 Push
the Zero Button.
3.5. Press
the PEAK button
3.6. Select the unit in which torque value
is required to display / measure. The torque value required for Blue Hawaii
hair oil is Ncm.
3.7 Apply
the force on the cap / seal to open it manually.
3.8 The
maximum torque value will be held, if peak was selected.
3.9 Note
the reading
3.10
To bring the system in normal
condition, push the PEAK button once again.
3.11 Repeat operation 3.2 to 3.10 for each and
every bottle checked.
4.0 FREQUENCY:
As and when required.
5.0 RESPONSIBILITY: Q.A
Executive.
6.0 REVIEW
GUIDE: Equipment’s manual.
7.0 ANNEXURE:
---------