STANDARD OPERATING PROCEDURE CALIBRATION OF POLARIMETER
STANDARD
OPERATING PROCEDURE
SUBJECT :
CALIBRATION
OF POLARIMETER
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Calibration of Polarimeter.
2.0 POLICY:
To ensure establish a procedure for Calibration of
Polarimeter.
3.0 PROCEDURE:
3.1
Put the switch on and leave the
equipment for warm up of lamp for 30minutes.
3.2
Prepare sucrose solution of the
concentration mentioned below.
3.3
Clean the polarimeter tube with
distilled water.
3.4
Check that the light source should
be properly aligned with respect to the optical bench.
3.5 Transfer
the solution to the polarimeter tube within 30min from the time the substance was
dissolved, during this interval maintains the solution at 25 deg. C.
Concentration
(g/100 ml) Angle
of Rotation (+) at 25 deg. C
10.0 13.33
20.0 26.61
30.0 39.86
40.0 53.06
50.0 66.23
4.0 FREQUENCY: Once in three
months.
5.0 RESPONSIBILITY:
Executive (QA).
6.0 REVIEW
GUIDE: Calibration register.
7.0 ANNEXURE:
-------
8.0 SOPs
DISTRIBUTION: