STANDARD OPERATING PROCEDURE CALIBRATION OF SHIMADZU HPLC (LC-2010CHT)
STANDARD
OPERATING PROCEDURE
SUBJECT:
CALIBRATION
OF SHIMADZU HPLC
(LC-2010CHT)
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
VERSION :
|
1.0 OBJECTIVE:
To provide guidelines for Calibration
of Shimadzu HPLC.
2.0 SCOPE:
To establish a procedure for
calibration of Shimadzu HPLC.
3.0 PROCEDURE:
3.1 Switch on the instrument and connect to LC solution software.
3.2 Press the Purge button, select the required port and
purge by using display panel present on instrument. Purging can also be performed
by gradually increasing the flow rate of the pump by using software.
3.3 Solvents
used are Methanol, IPA, and Water.
3.4. Calibration of the pump
3.4.1 Check the flow rate of the filtered and
degassed Purified water as follows.
3.4.2 Disconnect any column if connected to the
system and connect flow restrictor. Set the
flow of the
Channel “A” at 0.500 ml per minute. Keep the
system at this flow rate for 10 minutes to equilibrate the
system.
3.4.3 Set the flow to 1.000 ml per minute and 2.000ml per minute
respectively and repeat the above experiment, allowing the system to
equilibrate at least for 10 minutes at each flow set.
3.4.4 Repeat the steps for setting the flow using
channel “B, C and D”
3.4.5 Acceptance criteria:
Flow accuracy: Between 95% and 105% of the set flow rate.
5 Wavelength accuracy (Using Caffeine in purified water)
3.5.1 Caffeine Solution:
Set the following chromatographic
conditions.
Column
|
:
C18, 250 mm x 4.6 mm, 5 microns
|
Mobile
phase
|
:
Methanol: Purified water (40: 60 v/v). Filter and degas
|
Flow
rate
|
:
1.0 ml / min.
|
Wavelength
|
:
272 nm
|
Injection
volume
|
:
5, 10, 15, 20 µl
|
Column
oven temp
|
:
30°C
|
3.5.2 Preparation of standard
solution: Weigh accurately 100mg of caffeine, transfer into a 100 ml
volumetric flask, dissolve and dilute to volume with purified water. Further dilute 2 ml of this solution to 100 ml with
purified water. (Concentration 20ppm Caffeine)
3.5.3 Install the flow restrictor in the column thermostat,
equilibrate and keep sample solution in auto sampler.
3.5.4 Switch ‘ON’ the detector UV lamp for at
least 30 minutes.
3.5.5 Inject the Caffeine solution & record
the results.
3.5.6 Acceptance
Criteria:
Acceptable range: 270-276 mm.
3.6 Accuracy of auto injector and linearity
of UV- Detector response.
Chromatographic
conditions same as above:
3.6.1 Preparation of standard
solution: Weigh accurately 100mg of caffeine, transfer into a 100 ml
volumetric flask, dissolve and dilute to volume with purified water. Further dilute 2 ml of this solution to 100 ml with
purified water. (Concentration 20ppm Caffeine)
3.6.2 Theoretical plates of caffeine peak should not be less than 3000
and tailing factor should not be more than 2.0.
3.6.3 Calculate the % RSD of the peak areas & retention time of
caffeine from the 6 replicate injections at each level. Calculate the Mean peak
area of caffeine from the 6 replicate injections at each level.
3.6.4 Plot a linearity graph between injections volumes on the X-axis
V/S mean area of caffeine at each level on the Y-axis (instrument linearity
graph can be considered for the same).
3.6.5
Acceptance Criteria:
1) The % RSD of retention time of
caffeine from the six replicate injections at each level should be less than 2%
and % RSD of Area of caffeine from the six replicate injections at each level
should be less than 2%
2) The graph of caffeine for different levels
should be Linear.
3.7 Carry Over Test:
3.7.1 Chromatographic
conditions same as above:
3.7.2 Preparation of standard
solution: Weigh accurately 100mg of caffeine, transfer into a 100 ml
volumetric flask, dissolve and dilute to volume with purified water. Further dilute 2 ml of this solution to 100 ml with
purified water. (Concentration 20ppm Caffeine)
3.7.3 Inject blank followed by standard in
triplicate and again blank form fresh vial.
Calculate
any carryover of caffeine in the blank.
3.7.4 Calculate % carryover of Caffeine peak, in
the blank chromatogram after 20 µl of standard solutions,
by using formula:
Area of Caffeine in blank
=
------------------------------------------------ x 100
Mean
area of Caffeine standard
3.7.5 Acceptance
Criteria: Carry over should not be more than 0.1%.
3.8 Gradient composition:
3.8.1 Set the following chromatographic
conditions.
Column :
C18, 250 mm x 4.6 mm, 5 microns
Solvent
for channel A : Filtered and degassed purified water.
Solvent
for channel B / C / D : 0.1 % Acetone in purified water
Flow
rate : 2.0 ml/min
Wavelength
: 265 nm
Gradient
programming:
Step
|
Time in
minute
|
%
Solvent – A
|
%
Solvent – B / C / D
|
1
|
0.0
|
100.0
|
0
|
2
|
1.0
|
100.0
|
0.0
|
3
|
4.0
|
99.0
|
1.0
|
4
|
8.0
|
50.0
|
50.0
|
5
|
12.0
|
1.0
|
99.0
|
6
|
15.0
|
0.0
|
100.0
|
7
|
24.0
|
100.0
|
0.0
|
3.8.2 Procedure: Set the flow rate of at 2.0 ml per minute. Keep the
system in this flow rate for some time to equilibrate. Start the gradient
programming.
3.8.3 Acceptance
Criteria:
As per the fed gradient program the
system functioned.
Result:
The accuracy of Gradient composition passes as per the desired program.
3.8.4 The
instrument is covered under annual service contract and it is serviced as and
when there is a breakdown or requirement.
4.0 FREQUENCY: Internal Calibration:
Once in three months for flow rate and Gradient
composition. Once in six months for Wavelength accuracy, Accuracy of auto injector, linearity
& Carry Over test.
External
Calibration:
Once a year for wave length accuracy check flow rate
Accuracy, Injection volume accuracy Oven Temperature accuracy Repeatability and
check of program files by SPINCO LABORTORY PVT LTD Service Engineer.
5.0 RESPONSIBILITY: G.M (Q.A)/ Executive Q.C
Version
No : 13
6.0 REVIEW GUIDE:
Calibration record.
7.0 ANNEXURE:
--------
8.0 SOPs DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master
Copy
|
Circulatory
Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|
Maintenance
|
x
|
x
|
THNKS FOR PROVIDING THIS SOP
ReplyDeleteCAN you provide this types of sop formats for waters alliance 2695
Automation is crucial in pharmacy work, and PK Medi’s solutions drastically reduce the chance of human error. We’ve noticed a big improvement in safety.
ReplyDeleteFor More Details: https://pkmec.com/