STANDARD OPERATING PROCEDURE Operation and Calibration of Antibiotic Zone Reader
STANDARD OPERATING PROCEDURE |
||
Department: Quality Control
|
SOP No.: QC/MB/xxxx
|
|
Operation
and Calibration of Antibiotic Zone Reader
|
Effective
Date:
|
|
Review
Date:
|
||
Page
No.: xxxx
|
1.0
OBJECTIVE:
1.1
To lay down a procedure for Operation and Calibration of Antibiotic zone
reader.
2.0
SCOPE:
2.1
This Standard operating procedure is applicable for
operation of antibiotic zone reader
located in microbiology
laboratory at Pharma Private Limited.
I.D No.: xxx
Make: In
Lab Equipments.
3.0
RESPONSIBILITY:
3.1
Microbiologist – Responsible for operation of the
instrument.
3.2
Section In-charge – Responsible for review of the
procedure.
3.3
QC Head – Responsible for implementing the procedure.
3.4
QA Head – Overall compliance
4.0
PROCEDURE:
4.1
Operating Procedure:
4.1.1
Switch on the Antibiotic zone reader.
4.1.2
Turn the right side drum anticlockwise to its maximum
position, for the digital indicator to show zero reading or press the red
button for zero indication.
4.1.3
Place the Petri dish on the rotating plate, rotate the
petri dish manually, so that one end of the zone image just touches the mark
line on the prism.
4.1.4
Now turn the right side drum slowly, in clockwise
position till the other end of the zone image touches the mark line on the
prism.
4.1.5
Note the readings shown on digital indicator. The digital indication is
available from 0 to 35 mm.
4.1.6
Switch off the instrument when not in use.
4.1.7
Enter the usage details in Annexure – I (Antibiotic Zone Reader usage log
sheet).
4.2
Calibration Procedure:
4.2.1
Turn the right side drum anticlockwise to its maximum
position, for the digital indicator to show zero reading or press the red
button for zero indication.
4.2.2
Place the calibrated stainless steel /acrylic discs of
known diameter (18mm) in petri dish on the rotating plate, rotate the petri
dish manually, so that one end of stain less steel disc image just touches the
mark line on the prism.
4.2.3
Now turn the right side drum slowly, in clockwise
position till the other end of the stainless steel disc image touches the mark
line on the prism.
4.2.4
Note the readings shown on digital indicator in Annexure II.
4.2.5
Now reset the digital display reading to zero by pressing reset button.
4.2.6
Likewise measure the other three disc of each of different diameter
(20mm, 25mm and 27mm) and record the results in Annexure II “Calibration Report of Antibiotic Zone Reader”.
4.2.7
Acceptance Criteria:
4.2.7.1 The measured
diameter shall be within the ± 0.2mm of calibrated diameter.
4.2.8
If antibiotic zone reader is failed to meet an acceptance criteria of
calibration, then out of calibration shall be investigated as per current
version of SOP QA/GN/xxx.
4.2.9
Calibration frequency: Quarterly ± 7 days or after any major maintenance
work.
5.0
ABBREVIATIONS:
QC: Quality Control
|
QA: Quality Assurance
|
SOP:
Standard Operating Procedure
|
mm: Milli Meter
|