STANDARD OPERATING PROCEDURE Operation and Calibration of Analytical Balance
|
STANDARD OPERATING PROCEDURE |
|
Department: Quality Control
|
SOP No.: QC/MB/
|
|
Operation and Calibration
of Analytical Balance
|
Effective
Date:
|
|
Review
Date:
|
||
Page
No.:
|
1.0
OBJECTIVE:
1.1
To lay down a procedure for Operation, Calibration and cleaning for
Analytical Balance.
2.0
SCOPE:
2.1
This procedure is applicable to Analytical Balance located in
microbiology laboratory at Pharma Private Limited.
ID No.
Make: Mettler Toledo
Model No.:
Minimum weight: 0.2 g,
Maximum weight: 120 g,
‘d’ value: 1mg
Operational Range: 820 mg to 100 g
3.0
RESPONSIBILITY:
3.1
Microbiologist: Responsible for operation of instrument.
3.2
Section In-charge: Responsible for review the
procedure
3.3
QC Head: Responsible for implementing the procedure.
3.4
QA Head: Overall compliance.
4.0
PROCEDURE:
4.1
Precautions during measurement:
4.1.1
Do not use any organic solvents to clean the balance. (Ensure cleaning solution is not directly used).
4.1.2
Do not allow the material to remain on the balance for an extended period
of time.
4.1.3
The total weight on the pan of the balance
should not exceed the operating range.
4.1.4
In case of any defect immediately unplug the
instrument from mains.
4.2
Cleaning Procedure:
4.2.1
Clean the balance with nylon brush or lint free cloth.
4.2.2
Wipe down
draft shield doors, weigh pan and exterior with lint free cloth which is soaked
in mild detergent or approved disinfectant.
4.3
Operating procedure:
4.3.1
Ensure that the balance is clean and free from
dust.
4.3.2
Ensure
balance is on a solid, level surface, away from drafts and sources of
vibration.
4.3.3
Use
adjustable feet to adjust bubble level which is located at the front of the
unit to ensure the balance is level prior to use.
4.3.4
Plug in
unit (if necessary) and press the
key.
Balance should be allowed to “warm up” for at least 45 minutes prior to use. (To enable the balance to adapt itself to the
ambient conditions).

4.3.5
Long press
key,
the balance will show
followed by
this indicates the balance is in standby
mode.



4.3.6
Then press
and hold the
key then the display will show
.


4.3.7
Now the
balance is ready to weighing.
4.3.8
Press
key to
convert the measurement unit in ‘mg’.

4.3.9
Place the
sample and note down the weight appears on display.
4.3.10
After
weighing transfer the sample and then press
key to
tare it.

4.3.11
After usage, long press
key then display will show
followed by
, this indicates the balance is in standby
mode.



4.3.12
Always keep the balance in standby mode only.
4.3.13
If balance is in standby mode it is immediately
ready for weighing (warm-up period is not required).
4.3.14
Usage details of analytical balance shall be
entered in Annexure – III.
4.4
Calibration Procedure:
4.4.1
Daily Calibration:
4.4.1.1
Internal Calibration (Auto Calibration):
4.4.1.1.1
Press and hold
key until
appears.


4.4.1.1.2
Then select
by pressing
key.


4.4.1.1.3
The display will show
.

4.4.1.1.4
Then press
key to
execute “internal adjustment”.

4.4.1.1.5
The balance adjust itself automatically and finally the display will show
if the
adjustment done properly.

4.4.1.1.6
Internal Calibration (Auto calibration) shall be done when ever needed.
4.4.1.2
External Calibration (Daily weight
verification):
4.4.1.2.1
For weight verification, check the
weights of 1g, 10g, 50g & 100g and records the observed weights in
Annexure-I.
4.4.1.2.2
Weight verification shall be carried out at the starting of the day or
whenever the weighing activity is carried out.
4.4.1.2.3
Acceptance criteria: Each individual value should not
differ ± 0.1% of mass value.
4.4.2
Periodic Calibration:
4.4.2.1
Sensitivity:
4.4.2.1.1
Definition: Change in weighing value divided by
the change in load.
4.4.2.1.2
Place 100 g of the standard weight on the weighing pan, after the balance
get stable record the observed weights in Annexure –II.
4.4.2.1.3
Calculate the sensitivity of the balance by
using following formula
Sensitivity
= ((Mass value – Observed
value)/ Mass value) x 100
4.4.2.1.4
Acceptance Criteria: Should not be more than ±0.05%.
4.4.2.2
Linearity:
4.4.2.2.1
Definition: Ability of a balance to follow the
linear relationship between a load and the indicated weighing value.
4.4.2.2.2
Linearity Level A: Place 1 g of the standard weight on the weighing pan, after the balance
get stable record the observed weights in Annexure –II.
4.4.2.2.3
Linearity Level B: Place 10 g of the standard weight on the weighing pan and press the tare
key, then place 1g of standard weight on the weighing pan after the balance get
stable record the observed weights in Annexure –II.
4.4.2.2.4
Linearity Level C: Place 20 g of the standard weight on the weighing pan and press the tare
key, then place 1g of standard weight on the weighing pan after the balance get
stable record the observed weights in Annexure –II.
4.4.2.2.5
Linearity Level D: Place 50 g of the standard weight on the weighing pan and press the tare
key, then place 1g of standard weight on the weighing pan after the balance get
stable record the observed weights in Annexure –II.
4.4.2.2.6
Linearity Level E: Place 100 g of the standard weight on the weighing pan and press the tare
key, then place 1g of standard weight on the weighing pan after the balance get
stable record the observed weights in Annexure –II.
4.4.2.2.7
Calculate the linearity of the balance by using
following formula
Linearity = (Maximum difference/ Maximum
capacity) x 100
Maximum
difference is the maximum value which is obtained from B-A, C-A, D-A and E-A.
4.4.2.2.8
Acceptance Criteria: Should not be more than ±0.05%.
4.4.2.3
Eccentricity:
4.4.2.3.1
Definition: The
asymmetrical placement of the center of gravity of the load relative to the
load receiver.
4.4.2.3.2
Place 50 g of the standard weight on center (C) of the weighing pan, after balance
get stable record the observed weights in Annexure –II.
4.4.2.3.3
Same procedure is applied by placing the 50 g of the standard weight on Top
(T), Bottom (B), Right (R) and Left (L) sides of the weighing pan
individually.
4.4.2.3.4
Calculate the eccentricity of the balance by
using following formula
Eccentricity = (Maximum difference/ Centre of mass value) x 100
4.4.2.3.5
Maximum difference is the maximum value which is obtained from C- T, C- B, C- R and C- L.
4.4.2.3.6
Acceptance Criteria: Should not be more than ±0.05%.
4.4.2.4
Repeatability:
4.4.2.4.1
Definition: Ability of a weighing balance to display identical
measurement values for repeated weighing of the same objects under the same
conditions.
4.4.2.4.2
Place 1 g of the standard weight on the weighing pan, after the balance get stable record the observed
weights in Annexure –II.
4.4.2.4.3
Repeat the same procedure for remaining nine measurements.
4.4.2.4.4
Calculate the repeatability of the balance by
using following formula
Repeatability = 2X Standard Deviation____ X 100
Desired
smallest net weight
Where A is the nominal net weight (Standard weight in mg).
4.4.2.4.5
If the standard deviation obtained is less than 0.41d (where d is the scale interval i.e. 1 mg), replace the standard deviation with 0.41d. Desired smallest net weight is 1g.
4.4.2.4.6
Acceptance criteria: Not more than 0.10%.
4.4.2.5
Accuracy:
4.4.2.5.1
Definition: The degree to which the result of a
measurement, confirms to the correct value or a standard.
4.4.2.5.2
Place 20 g of the standard weight on the weighing pan, after balance get
stable record the observed weight in Annexure –II.
4.4.2.5.3
Repeat the same procedure for remaining nine measurements.
4.4.2.5.4
Calculate the accuracy of the balance by using
following formula
Accuracy = ((Mass value – Mean value)/ Mass value) x
100
4.4.2.5.5
Acceptance criteria: Not more than 0.10%.
4.4.2.6
Weighing Performance Test:
4.4.2.6.1
Definition: To
check the performance of the balance throughout the range of the balance.
4.4.2.6.2
Place the respective standard weights (1g, 2g, 5g, 10g, 20g, 50g and
100g) on the weighing pan individually after balance get stable record the each
observed weight in Annexure II.
4.4.2.6.3
Acceptance Criteria: Should not be more than ±0.10% of mass value.
4.4.2.7
Periodic Calibration frequency: Once in a month ± 3 days or after any major maintenance
work.
4.4.3
If analytical balance is failed to meet an acceptance criteria of
calibration, then out of calibration shall be investigated as per current
version
5.0
ABBREVIATIONS:
QC: Quality Control
|
QA: Quality Assurance
|
UPS: Uninterrupted
Power Supply
|
Mg: Milli gram
|
SOP: Standard Operating Procedure
|
g : Gram
|
Comments
Post a Comment