STANDARD OPERATING PROCEDURE LAB EQUIPMENT AUTOMATIC DIGITAL REFRACTOMETER MODEL: DG-NXT
STANDARD
OPERATING PROCEDURE
SUBJECT :
LAB EQUIPMENT AUTOMATIC DIGITAL REFRACTOMETER MODEL:
DG-NXT
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Operation & Maintenance
of Automatic Digital Refractometer.
2.0 POLICY:
To establish a procedure for an operation of Automatic Digital Refractometer.
3.0 PROCEDURE:
3.1. Open the Flap cover of the prism and ensure
the prism surface is free from dust, or any other particle. Incase not then
clean the prism surface gently with distilled water and wipe the same with
Cotton or Tissue paper
3.2. Switch on the instrument by pressing ON / OFF
switch. The current power level of the 9V battery fitted is shown on the top
right corner of the LCD screen.
3.3 Select the desired scale (Refractive Index, nD
/ Brix) by pressing the scale key & keep it pressed till the scale indication
points on the desired scale shown at the bottom of the screen for P&G Blue
Hawaii select the refractive index option.
3.4. Calibrate
the instrument by using DI water. Place approx 1 ml of DI Water on the prism
surface. Close the flap cover, wait for 30 seconds. Calibrate the instrument by
pressing the Zero key pressed till CAL is displayed on the screen. Then release
the calibration reading is displayed on the screen, RI of water is 1.3333. Then
pres zero key. The display will be ……
3.5. In case
Temperature compensated reading is required then the automatic Temperature
compensation function needs to be switched ON (Indicated by the symbol “ATC” on
the LCD screen). In case “ATC” is not appearing on the screen press FUNCTION
key thrice. Then “ATC.Y” will appear on the screen. Select the option Y by
pressing the scale key. Then press Read/Enter key to confirm your setting. Now
recheck whether ATC appearing on screen. If this function needs to be switched
off, same procedure needs to be followed wherein after pressing Function key
thrice. Then “ATC.N” will appear on the screen which needs to be select.
3.6. Take small amount of sample (approx
1ml) to be measured with the help of a dropper and put in on the prism surface
gently. Close the flap cover. Press the READ key. The value will instantly be
displayed on the LCD screen. There after the measurement is done gently clean
and wipe the prism with distilled water and tissue paper after use.
4.0 FREQUENCY:
As and when required.
5.0 RESPONSIBILITY: QA
Executive
6.0 REVIEW
GUIDE: Equipment’s manual.
7.0 ANNEXURE: ---------