STANDARD OPERATING PROCEDURE LAB EQUIPMENT - SINGLE PAN ANALYTICAL BALANCE (Make: AFCOSET & CITIZEN)
STANDARD
OPERATING PROCEDURE
SUBJECT :
LAB EQUIPMENT - SINGLE PAN
ANALYTICAL BALANCE
(Make: AFCOSET &
CITIZEN)
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Operation & Maintenance
of Single Pan Balance.
2.0 POLICY:
Balance is the heart of the system .Its accuracy is of
vital importance in the analysis so it should be handled in accordance with the
SOP.
3.0 PROCEDURE:
3.1
Check the spirit level and ensure
that the balance is in stable position.
3.2
To switch on the balance connect the
plug to the main and put the switch on.
3.3
Press the ON/OFF button. The display
will be 0.0000g.
3.4 To
weigh place a piece of butter paper on the pan close the side doors and wait
until the display is stable.
3.5
The displayed reading is the weight
of the paper.
3.6
To tare the weight press” Tare”
knob. The weight gets tarred and the display will be 0.0000g.
Place the object to be weighed on the butter paper and
close the doors. Wait until the display reading is stable. The display reading
is the weight of the object.
3.6.1
Repeat the same
operations from step 3.4 to step 3.6 for further weighing.
1.8.
When weighing is
done for Assay purpose do not tare the weight of butter paper instead record
the weight of butter paper and then add the desired quantity of sample and
determine the weight. Record the weight of sample with butter paper. Transfer
the sample into the required glassware for testing and reweigh the butter
paper. The difference in weight is the sample taken for Assay. This need to be
done while weighing standards and samples for Assay testing so as to reduce the
error if any at lower weight ranges.
3.5
Afcoset balance is used for weighing
regents for preparation of volumetric solutions, test solutions
and other
preparations used for analysis.
3.10 MAINTAINENCE
3.9.1
When the work is
over switch off the balance.
3.9.2
Always keep the
pan dust free and dry, clean it with small brush provided if necessary.
3.9.3
Always keep the
side glass doors closed while taking the readings to get accurate weight of the
object and also keep it closed while it is not in use to avoid dust accumulation
on the pan.
3.9.4
Always keep the
balance away from vibrations.
3.9.5
Balance must not
be over loaded.
3.9.6
Hot objects
should be cooled to room temperature before putting on the balance pan.
3.9.7
Should any
spillage occur during the weighing it should be cleaned immediately?
3.9.8
A stoppered
container must be used for weighing volatile liquids or hygroscopic solids.
3.10.9
Switch off the balance and clean the pan
and platform of the balance with tissue paper frequently
after weighing.
4.0 FREQUENCY: As and when
required. Calibration once in three months.
5.0 RESPONSIBILITY:
Executive Q.A.
6.0 REVIEW
GUIDE: Calibration record.
7.0 ANNEXURE:
-------
8.0 SOPs
DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|
Maintenance
|
x
|
x
|