STANDARD OPERATING PROCEDURE OPERATION AND MAINTENANCE OF DISSOLUTION TEST APPARATUS (TDT-06 P)
STANDARD
OPERATING PROCEDURE
SUBJECT :
OPERATION AND MAINTENANCE OF DISSOLUTION TEST APPARATUS(TDT-06 P) |
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for operation and Maintenance of
Dissolution Test apparatus.
2.0 POLICY:
Dissolution test apparatus is used for testing the
dissolution profile of tablets so it is necessary to maintain it in orderly
condition and to check its performance.
3.0 PROCEDURE:
3.1 Switch
on the power.
3.2 After the power is on the instrument will be initialized and
the display will show start indicating the manual mode of operation.
3.3 Switch
on the Heat switch provided on the rear right hand side of the stirrer unit.
3.4 After
the power is switched the following are displayed.
3.4.1 The STEP and TIMER display shows “Start”.
3.4.2 The RPM display shows “000”.
3.4.3 The temperature display shows the present
temperature of the water bath.
3.5 Lift
the stirrer unit using the LIFT SWITCH provided at the rear right hand side.
3.6 Place
the test jars filled with the medium in the bath.
3.7 Fix the
paddle or baskets as desired for the test.
3.8 Now set the desired temperature. Press and hold the push
button switch and turn the set temperature knob in clockwise / counter
clockwise direction and lock the knob after setting is completed.
3.9 Now set the desired RPM by turning the SET RPM knob in
clockwise / counter clockwise direction and lock it. Now set the time parameter
by the computerized panel.
3.10 Test time is the cumulative time of all steps programmed in that
test. While setting the step timing makes sure that the current step time is
more than the previous step time. If the step timing is not entered correctly
then the tm-Err will be displayed.
3.11 To set time enter step and set the time in minutes and hours
with the help of arrow switches to increase and decrease. After the time is set
check the temperature of the medium in the bowls.
3.12 Place
one tablet inside the test bowl.
3.13 Bring
the stirrer unit down using the lift switch. The stirrer unit will
automatically stop.
3.14 Start
the test by pressing the Run / Halt key.
3.15 At the
end of the run time the buzzer beeps for a few seconds.
3.16 At the end of the specified time collect the medium from each
bowl and test as per the specified test procedure.
3.17 After
completion of testing ensure that the bowls and paddles are cleaned thoroughly.
3.18 CALIBRATION:
Check the temperature of the water bath with a calibrated thermometer and check
the RPM of the paddle once in 3 months.
4.0 FREQUENCY
OF USE: Use as and when required.
5.0 RESPONSIBILITY:
Executive QA & Chemist QA.
6.0 REVIEW
GUIDE: Calibration register.
7.0 ANNEXURE:
-------
8.0 SOPs
DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|
Maintenance
|
x
|
x
|
Comments