STANDARD OPERATING PROCEDURE OPERATION AND CALIBRATION OF BROOK FIELD VISCOMETER
STANDARD
OPERATING PROCEDURE
SUBJECT:
OPERATION AND CALIBRATION
OF BROOK FIELD VISCOMETER
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To
provide guidelines for Operation & Calibration of Brookfield Viscometer.
2.0 POLICY:
To
determine the viscosity of viscous nature of the various Liquids, Creams and
oils etc.
3.0 PROCEDURE:
3.1
Switch on the Viscometer.
3.2
Select the Spindle number or speed by
adjusting the select knob on either side by adjusting the
round knob in the middle as per the
Specifications.
3.3
Set the RPM as
per the Specifications, by rotating the round knob in the middle of the Viscometer.
After setting the RPM &
Spindle number, bring the select switch settings to normal position.
3.4
Fix the disc spindle or helipath spindle as
required to viscometer, Immerse the spindle
into the sample contained in a 500ml beaker until the
mark on the spindle rod adjust the
height. If
Helipath spindle is used it is fixed by helipath assembly to viscometer and the
spindle will
pierce through the cream. Incase of helipath spindle the spindle need not be
immersed in the
cream. Adjust the helipath spindle with the flat knob on the helipath
stand.
3.5
Switch ON the button, the spindle starts
rotating as per the set parameters.
3.6
Note the reading
when the displayed value is stable.
3.7
Remove the
spindle, clean and place in spindle box. Switch off the viscometer after use.
3.8
Enter the details
in the log book.
3.9 CALIBRATION:
3.9.1 The calibration and regular maintenance of the viscometer is done
by an external agency and is
covered
under Annual Maintenance contract.
3.9.2 The authorized service engineer of the external agency will check
the performance of the
viscometer
and calibrate the same with traceable standards available with them.
3.9.3 The agency will provide a calibration certificate which is valid
for a period of 6 months.
4.0 FREQUENCY
: As and when required.
5.0 RESPONSIBILITY :
Q.A Executive/chemist
6.0 REVIEW
GUIDE : Usage log book.
7.0 ANNEXURE:
-------
8.0 SOPs
DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|
Maintenance
|
x
|
x
|
Comments
Post a Comment