STANDARD OPERATING PROCEDURE CALIBRATION OF ANALYTICAL BALANCE (Make AFCOSET, CONTECH & SNUG, CITIZEN.)
STANDARD
OPERATING PROCEDURE
SUBJECT:
CALIBRATION OF ANALYTICAL
BALANCE (Make AFCOSET, CONTECH & SNUG, CITIZEN.)
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Calibration of Analytical
Balance.
2.0 POLICY:
Periodically Calibrated balances will be used for
maintaining Accuracy of balances.
3.0 PROCEDURE:
3.1 Switch
on the instrument the balance is automatically calibrated (citizen only)
3.2 The
Balance should be checked for leveling of the base by spirit level, before
operating. This
is applicable for Afcoset, Snug &
Citizen balances only.
3.3 The
level of contech balance is checked by placing spirit level on the balance
before operating.
3.4 Use Standard weights for the calibration
of the Balance. The standard weights are calibrated
and
tolerance limits are noted.
3.5
Place the Standard weights and note
the readings
3.6
Any difference in the Reading and
the Standard weight shall not be more than the
tolerance
limit as specified.
3.7
Daily check on higher weights- 1g,
2g, 5g, 10 g, 50g Lower weights – 20mg, 50mg, & 100mg.
3.8 All the readings shall be entered in the
Calibration register
3.9 The Afcoset Balance & Citizen
balances having a inbuilt Calibration facility which can be
used
for Calibration.
4.0 FREQUENCY: check daily
before use & Calibrate once in 3 months
5.0 RESPONSIBILITY:
Executive (QA)
6.0 REVIEW
GUIDE: Calibration Record.
7.0 ANNEXURE: -------
8.0 SOPs
DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|
Maintenance
|
x
|
x
|