STANDARD OPERATING PROCEDURE CALIBRATION OF EQUIPMENT
STANDARD
OPERATING PROCEDURE
SUBJECT :
CALIBRATION
OF EQUIPMENT
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To lay down a procedure with respect to frequency of
calibration of various Equipment and Maintenance of Records thereof.
2.0 POLICY:
Periodically calibrated Equipment will be used for
maintaining accuracy in analytical / dispensing / manufacturing process.
3.0 PROCEDURE:
3.1 The
equipment required to be calibrated is first listed and establish the
calibration procedure & frequency of calibration is established.
3.2 Calibration
procedures are prepared for individual equipment from the pharmacopoeias or
instruction manuals of equipments/in-house specifications.
3.3 Calibrate
different equipments according to their calibration procedure and frequency to
ascertain whether the performance of the equipment is satisfactory and
accurate.
3.4
All the working standards used for
calibration purpose are stored in a separate rack.
3.5 Results
obtained from instrument calibration is checked against standard results and
records are maintained.
3.6 Instruments
like H.P.L.C, UV-Visible spectrophotometer, Melting point apparatus,
Polarimeter, Vacuum Oven are calibrated once in every three months.
3.7
Analytical Balances, pH meter are
crosschecked daily before use.
3.8 Balances
used in Production and Dispensing area are crosschecked every day against
standard weights given by Govt. authorities and calibrated once in three months
by Q.A dept. and once a year by outside agency.
3.9 Calibrated
equipment should have a Calibration tag with the following details;
Name of equipment Calibration
due
Calibrated on Calibrated
by
Remarks
When any equipment if found Out of Calibration, report
to the Service Engineer for rectification. The equipment is to be Re-calibrated
after servicing is done. The calibration certificates received from external agencies
should be reviewed by Q.A and attested.
4.0 FREQUENCY: As and when required.
5.0 RESPONSIBILITY:
Executive (Q.A).
1.0
REVIEW GUIDE:
Calibration records.
7.0 ANNEXURE:
1. Calibration of equipment.
2. Standard Weights, Filling Machines Nozzles and
Weighing Scales: - External Calibration.
8.0 SOPs DISTRIBUTION:
This
SOP shall be distributed to following Departments:
Name of the Department
|
Master Copy
|
Circulatory Copy
|
QA/QC
|
√
|
√
|
Production
|
x
|
x
|
Warehouse
|
x
|
x
|
General
|
x
|
x
|