STANDARD OPERATING PROCEDURE CALIBRATION OF pH METER
STANDARD
OPERATING PROCEDURE
SUBJECT:
CALIBRATION
OF pH METER
|
DEPARTMENT : QA
SOP
NO. :
Page
No :
Effective
Date :
Next
Review :
SOP
REVISION :
|
1.0 OBJECTIVE:
To provide guidelines for Calibration of the pH meter.
2.0 POLICY:
To lay down a procedure for Calibration of PH meter.
1.0
PROCEDURE:
3.1 To
standardize the pH meter select two buffer solutions whose difference in pH
does not exceed four units, such that the expected pH of the material undertest
false between them.
3.2 Prepare the buffer solutions having pH 4, 7 & 9.20
3.2.1
Buffer pH 4.0
(IP-2007 4.1.B)
Place 2.86 ml of glacial aceticacid & 1.0 ml of
50% wt/volume solution of NaOH in 1000ml volumetric flask add water to volume
and mix. Adjusts the pH if necessary to 4.0.
3.2.2
Buffer solution
pH 7.0 & 9.2 are prepared as directed on the instructions of the pack.
3.2.3
Prepare the
buffer solutions on the first working day of the weak.
3.3 Switch
“on” the equipment. Remove the
electrode from the water; wipe out the water on the electrode using tissue
paper than rinse the electrode with CO2 free water than wipe out
with tissue paper.
3.4 Immerse
the electrode in one of the buffer solution, adjust the temperature control and
adjust the calibration control to the observed pH value identical with the
buffer solution under examination, by shaking it slowly.
3.5 Takeout the electrode washed with CO2
free water, wipe it with tissue paper.
3.5
Repeat the procedure using second buffer
solution.
3.7 Place
the electrode the third buffer solution and check the pH without adjusting the
calibration control at same temperature.
3.8 Note
down the pH readings of the solutions in Calibration register before making necessary
adjustment.
3.9
The readings of all three buffers should
not differ more than 0.05 units.
3.10 If the
difference is more than 0.05 units examine the electrodes & if they are
faulty, replace the electrode.
4.0 FREQUENCY: Daily.
5.0 RESPONSIBILITY: Executive (QA) / Jr. Executive (QA).
6.0 REVIEW
GUIDE: Calibration Record.